Power Density Measurement Technologies for Assessing Human Exposure to Electromagnetic Fields in the 5G and 6G Era
- Published: 2026-10-08
Power Density Measurement Technologies
Integrated Test Infrastructure Strategy for High-Frequency Human Exposure Compliance Assessment
1. Executive Summary
Potential Grabs
As the use of high-frequency wireless technologies such as mmWave, array antennas, and dynamic beamforming expands in 5G-Advanced and future 6G environments, human exposure assessment for mobile devices is evolving to consider not only conventional Specific Absorption Rate (SAR) measurements but also Power Density measurements. In particular, in the high-frequency range above 6 GHz, Power Density-based assessment is becoming increasingly important because RF energy tends to be concentrated near the skin and body surface rather than penetrating deeply into the human body.
High-frequency human exposure assessment is not a process of simply replacing SAR with Power Density. Appropriate assessment metrics and test methods must be applied according to the operating frequency, product type, and regulatory requirements, and the distinction between Incident Power Density and Absorbed Power Density must also be clearly understood. IEC/IEEE 63195-1:2022 and IEC/IEEE 63195-2:2022 serve as key standards for Power Density measurement and computation for wireless devices operating in close proximity to the head and body. IEC/IEEE 63195-1 addresses Power Density measurement procedures for wireless devices operating in close proximity over the frequency range of 6 GHz to 300 GHz, while IEC/IEEE 63195-2 addresses computational procedures for the same frequency range.
The accuracy and reproducibility of test results cannot be ensured solely by the specifications of individual measurement instruments. A test environment with RF shielding and reflection control, the spatial resolution of the Probe and Scanner, the positioning accuracy of the Positioner, calibration and system validation, and measurement uncertainty management must all be integrated into a single test workflow.
An efficient human exposure assessment infrastructure must go beyond simply installing a chamber and measurement equipment. Scanner and Positioner control, DUT operating modes and beam control, data acquisition, spatial averaging analysis, uncertainty estimation, automated report generation, and test history management must be integrated to reliably support product development, certification preparation, repeated testing, and quality control.
Problems and Solutions
Problems
- Conventional SAR measurement alone cannot fully cover the Power Density-based human exposure assessment required in the high-frequency range above 6 GHz. For DUTs using array antennas and beamforming, beam direction and power distribution, distance from the evaluation surface, and near-field electromagnetic field distribution can also affect measurement results.
- When multiple uncertainty factors such as Probe performance, positioning errors, chamber reflections, DUT alignment, and output drift overlap, the reliability and repeatability of results can be degraded. Calibration of individual equipment alone cannot ensure the measurement quality of the entire system.
- When control programs and data formats are separated by equipment, repeated testing takes longer and dependence on individual operators increases. As the number of test items and product categories expands, the risks of data omission and operational errors also increase.
- If a test laboratory is designed only around current products, it may be difficult to accommodate future 5G-Advanced and 6G candidate bands and new antenna architectures.
Solutions
- Assessment metrics and test procedures appropriate for the product frequency and operating conditions must be defined. Based on the IEC/IEEE 63195 series, a Power Density test plan should be established that reflects the evaluation surface, spatial averaging, DUT operating modes, and maximum exposure conditions.
- An architecture that enables integrated validation of the test environment and measurement system is required. Shielding and absorber design, precision Scanners and Positioners, calibration of Probes and measurement instruments, system validation, automated data analysis, and reporting systems should be implemented together.
- An automation platform that connects control, measurement, analysis, and reporting should be established. Scanner and Positioner control, DUT output and beam settings, measurement data acquisition, spatial averaging and maximum-value analysis, uncertainty incorporation, report generation, and database management can be integrated step by step.
- Modular measurement systems, replaceable Probes and Positioners, upgradeable software, and an expandable test environment should be adopted to accommodate new frequencies and assessment requirements.
Executive Summary Conclusion
Human exposure compliance assessment in the 5G and 6G era is not simply a matter of introducing a single measurement instrument. It is a challenge of building integrated test infrastructure that connects the test environment with measurement, positioning control, calibration, uncertainty management, automation, and reporting. Test laboratories and manufacturers that establish this infrastructure proactively can improve test reproducibility and operational efficiency throughout the development and certification processes for high-frequency wireless devices.
2. Why Human Exposure Testing Is Changing
Compared with previous generations of mobile communications, 5G-Advanced and next-generation 6G wireless communications are evolving toward the use of higher frequencies and wider bandwidths, while some high-frequency systems are adopting technologies such as array antennas, Massive MIMO, and dynamic beamforming. These technological changes affect not only communication performance but also require changes in the methods used to measure and assess human exposure to RF electromagnetic fields.
Specific Absorption Rate (SAR) has traditionally been used as a key measurement metric for assessing human exposure from hand-held and body-mounted wireless devices. SAR represents the power absorbed per unit mass of human tissue exposed to RF electromagnetic fields. IEC/IEEE 62209-1528:2020 specifies human models, measurement systems, and measurement procedures for SAR assessment of hand-held and body-mounted wireless communication devices operating from 4 MHz to 10 GHz.
However, as the use of high-frequency bands, including mmWave, expands in 5G NR FR2 and some future 6G candidate technologies, the measurement metrics and methods that need to be considered in human exposure assessment are also expanding. In the high-frequency range above 6 GHz, RF energy tends to be concentrated near the skin and body surface rather than penetrating deeply into the human body. Consequently, Power Density, which evaluates power per unit area from a different perspective than SAR, which represents power absorbed per unit mass, becomes an important assessment metric.
In human exposure assessment, Power Density can be distinguished into concepts such as Incident Power Density (IPD) and Absorbed Power Density (APD). Incident Power Density is an external exposure metric representing the energy flow of electromagnetic fields incident on the human body per unit area, whereas Absorbed Power Density is an internal exposure metric representing the power absorbed into tissue through the body surface per unit area. The two metrics therefore cannot be treated as the same physical quantity and must be distinguished according to the applicable exposure limits and assessment methods.
The IEC/IEEE 63195 series is currently used as an important international standard for high-frequency Power Density assessment of wireless devices operating in close proximity to the head and body.
- IEC/IEEE 62209-1528:2020
- SAR measurement procedures, human models, and instrumentation requirements for hand-held and body-mounted wireless devices from 4 MHz to 10 GHz
- IEC/IEEE 63195-1:2022
- Power Density measurement procedures for wireless devices operating in close proximity from 6 GHz to 300 GHz
- IEC/IEEE 63195-2:2022
- Power Density computational procedures for wireless devices operating in close proximity from 6 GHz to 300 GHz
The official scope of IEC/IEEE 63195-1 covers measurement procedures for assessing Power Density of human exposure to RF fields from wireless devices operating in close proximity over the frequency range of 6 GHz to 300 GHz. IEC/IEEE 63195-2 specifies computational procedures for the same type of devices and frequency range.
Changes in test methods are not limited to the assessment metrics. The antenna architectures and operating methods of the Device Under Test (DUT) are also becoming more complex. In high-frequency systems such as 5G NR FR2, the use of array antennas and beamforming can cause the spatial distribution of electromagnetic fields to vary depending on antenna beam direction and power distribution. Studies of Power Density compliance assessment for mmWave handsets have also identified that conventional planar near-field scanning can require significantly longer test times when multiple beam realizations need to be evaluated.
Therefore, high-frequency human exposure assessment must consider the distance between the DUT and the evaluation surface, measurement area and spatial resolution, Probe characteristics, DUT positioning and alignment, antenna beam conditions, and output conditions. To implement these requirements reliably, an integrated test system combining a test environment with RF shielding and reflection control, Probes, Scanners, Positioners, measurement systems, control software, and data analysis capabilities is required.
The changes in human exposure assessment can be summarized in three key points:
- First, human exposure measurement for mobile devices is expanding beyond conventional SAR measurement to include Power Density measurement in the high-frequency range.
- Second, for high-frequency wireless devices using array antennas and beamforming, a measurement system that considers evaluation distance, measurement position, spatial averaging, beam direction, and DUT operating conditions is becoming increasingly important.
- Third, ensuring accuracy and repeatability requires an approach that integrates not only individual measurement equipment but also the test environment, Probes, Scanners, Positioners, calibration and validation, automation, and data analysis into a single system.
Accordingly, test infrastructure for human exposure assessment in the 5G and 6G era should be designed as integrated Test Infrastructure that brings together an RF-shielded chamber and RF absorbers, precision Positioners, Scanners, Probes, RF measurement instruments, test automation software, and data management systems. This provides the foundation for ensuring the accuracy and reproducibility of high-frequency Power Density measurements and for accommodating evolving products and test requirements in the future.
3. Key Measurement Challenges
What matters in high-frequency Power Density measurement is not simply reading the strength of an electromagnetic field, but measuring the spatial electromagnetic field distribution generated by the DUT with sufficient accuracy and repeatability, spatially averaging it over an appropriate evaluation surface, and determining the maximum exposure condition. IEC/IEEE 63195-1:2022 also defines a Power Density measurement system as an integrated structure consisting of a Probe, Scanning System, Measurement Instrumentation, Post-processing, and Uncertainty Evaluation. In addition to the accuracy of the measurement system itself, environmental and operational variables such as DUT alignment, surrounding reflections, measurement distance, and device drift are also treated as uncertainty contributors.
Particularly in the mmWave range, wavelengths are short and electromagnetic field distributions can change rapidly in the near field. As a result, positioning errors or spatial sampling intervals that may have been acceptable in lower-frequency testing can have a greater impact on measurement results. IEC/IEEE 63195-1 separately identifies Probe Positioning Offset, Probe Positioning Repeatability, Probe Spatial Resolution, Field Reconstruction, Spatial Averaging, DUT Alignment, and Laboratory Reflection as independent contributors to measurement uncertainty. Therefore, when designing a test system, what matters more than the specifications of individual instruments is how effectively these variables can be controlled and validated across the entire measurement system.
The major variables that should typically be considered in Power Density measurement can be summarized as follows.
| Key Measurement Challenge | Impact on Measurement Results | Considerations for Test Infrastructure Design |
|---|---|---|
| Probe Accuracy & Calibration | Probe sensitivity, frequency response, isotropy, linearity, and Cross Coupling directly affect measured values | Select a Probe suitable for the operating frequency and signal characteristics, and manage calibration history and calibration uncertainty |
| Spatial Resolution & Scan Area | If the measurement interval is too large or the measurement area is insufficient, localized peak values may be missed | Secure an appropriate Grid interval for the wavelength and measurement method, as well as a sufficient Scan Area |
| DUT–Evaluation Surface Distance | In the near field, E/H-field and PD distributions may vary significantly with distance | A Positioning structure capable of accurately reproducing the distance between the DUT and the evaluation surface is required |
| DUT Positioning & Alignment | Position and angular errors can affect the location of maximum exposure and the psPD result | Establish repeatable DUT Fixtures, Positioners, and Alignment reference points |
| Beam & Output Condition | For beamforming DUTs, the maximum exposure location can vary depending on beam direction and output conditions | Define the Beam and Operating Mode to be tested, identify the maximum exposure condition, and stabilize output |
| Spatial Averaging | Even with the same Field Distribution, psPD may vary depending on Averaging Area and processing method | Apply spatial averaging and maximum-finding algorithms in accordance with the applicable requirements |
| Reflection & Ambient RF | Reflections from chamber walls, fixtures, and surrounding structures can be superimposed on the measured electromagnetic field | Apply RF shielding and reflection control, select appropriate DUT Holder materials, and validate the test environment |
| Measurement Uncertainty | Errors from the Probe, Positioning, Sampling, Reconstruction, and DUT Drift can accumulate | Manage the uncertainty Budget based on the entire Measurement Chain rather than individual errors |
| Repeatability & Reproducibility | Variations may occur between repeated tests or when the operator or laboratory changes | Manage System Check, System Validation, test conditions, and Calibration history |
Probe Accuracy and Calibration
In Power Density measurement, the Probe is a critical measurement element that converts the electromagnetic field around the DUT into actual measurement data. The accuracy of the Probe is therefore directly linked to the reliability of the overall test result.
IEC/IEEE 63195-1 requires consideration of characteristics related to the Field Probe, including Sensitivity, Frequency Response, Isotropy, Linearity, Sensor Cross Coupling, Probe Scattering, Mechanical Offset, and Field Impedance Dependence. In particular, the Probe Calibration Report should include the uncertainty and applicable conditions associated with these characteristics, and variations in characteristics according to frequency and bandwidth should also be managed.
In high-frequency near-field measurements, the physical size of the Probe itself also becomes an important factor. In regions where the electromagnetic field varies rapidly in space, the Probe does not measure the electromagnetic field at an ideal single point, but may instead exhibit a response over a certain physical volume. IEC/IEEE 63195-1 therefore manages this difference separately as uncertainties associated with Probe Correction and Probe Spatial Resolution.
Accordingly, it is not sufficient for a test system to verify only a frequency range specification such as “a Probe capable of measurements up to 300 GHz.” The Probe and measurement system should be selected by considering the intended measurement distance and Field Gradient, required spatial resolution, the E-field or H-field components to be measured, whether Phase information is required, and the Dynamic Range.
Scanner Spatial Resolution and Measurement Area
Power Density assessment measures the electromagnetic field distribution over an evaluation surface and uses this data to calculate Spatial-average Power Density and Peak Spatial-average Power Density (psPD). The ability of the Scanner to establish measurement points with sufficient spatial resolution and positioning accuracy therefore directly affects the reliability of the results.
According to IEC/IEEE 63195-1:2022, when Power Density is evaluated by directly measuring the E-field or H-field on the Evaluation Surface, a Step Size of λ/4 or less is required in the Near-field, while in the Far-field it may be relaxed, for example, to λ/2 or less.1 For measurements using Forward or Backward Transformation, the same λ/4 condition is not applied uniformly. Instead, the required Scan Area and Spatial Resolution should be separately defined, validated, and documented according to the measurement system and Field characteristics.2 In particular, higher spatial resolution may be required in close-proximity regions where the influence of Evanescent Modes is significant.
As frequency increases, λ becomes shorter, which may require a greater number of measurement points at smaller intervals even when measuring the same physical area. This directly affects not only the mechanical positioning accuracy of the Scanner but also the overall test time.
If the Scan Area is too small, the region containing the actual maximum Power Density may fall outside the measurement range, or insufficient spatial information may be available for Field Reconstruction. The Scanner should therefore be designed not simply as a device that moves precisely, but as a Positioning System capable of repeatedly measuring a sufficient area at the required resolution.
1 IEC/IEEE 63195-1:2022, Clause 7.4.2.2(b): For the direct measurement method, the measurement step size in the near field is required to be λ/4 or less, while in the far field it may be relaxed, for example, to λ/2 or less. ↩
2 IEC/IEEE 63195-1:2022, Clause 7.4.2.3(b): For forward/backward transformation methods, the scan area and spatial resolution should be defined, validated, and documented according to the measurement system and field characteristics, and higher spatial resolution may be required when evanescent modes are significant. ↩
Distance Between the DUT and the Evaluation Surface
Power Density assessment for high-frequency wireless devices operating in close proximity is performed with reference to an Evaluation Surface located at a specified distance from the DUT. IEC/IEEE 63195-1 defines the Evaluation Distance as the distance between the DUT and the Evaluation Surface. When the Measurement Region and the actual Evaluation Surface are different, the measurement data may be reconstructed to determine the Power Density on the evaluation surface.
This distance is not merely a mechanical dimension. Particularly in the Near-field, a change in distance from the DUT can alter not only the magnitude of the electromagnetic field but also the relationship between the E-field and H-field and the spatial field distribution itself. Consequently, Positioning Errors on the order of several millimeters can affect the measurement result, and the distance between the DUT and the evaluation surface must be maintained consistently when the same test is repeated.
For this purpose, reference points should be clearly defined across the entire mechanical structure, including the DUT Holder and Positioner, so that the relative position between the Probe and the DUT can be reproduced repeatedly. Rather than evaluating a Positioner solely by its travel range or number of axes, it is important to design it based on the accuracy and repeatability of the relative coordinates among the DUT, Probe, and Evaluation Surface.
DUT Positioning and Alignment
For DUTs using array antennas and Beamforming, even small positional or rotational errors can affect the measured Field Distribution and peak location. IEC/IEEE 63195-1 also identifies DUT Alignment as one of the environmental and DUT-dependent contributors to measurement uncertainty.
What matters in testing is not simply positioning the DUT accurately once, but whether it can be installed repeatedly according to the same reference. Fixtures and Reference Points should be defined based on the product geometry and antenna location, and the Positioning Architecture, including Translation and Rotation, should be capable of reproducing the same test position.
In particular, when evaluating multiple antenna modules or different Device Orientations, it is advantageous to minimize manual alignment by the operator and to use a structure in which Position and Orientation can be recorded and recalled through software. This also directly supports future automation and test history management.
Beam Direction and DUT Output Condition
Array antennas and Beamforming used in 5G NR FR2 handsets do not have a single fixed Radiation Pattern. Instead, the spatial distribution of the electromagnetic field can vary according to the Beam Configuration. Therefore, simply operating the DUT at maximum output does not necessarily ensure that the maximum exposure condition has been identified.
Studies of PD assessment for actual 5G mmWave handsets have indicated that evaluating multiple Beam Realizations individually using conventional Planar Near-field Scanning can require substantial test time. For products with dozens or hundreds of Beam Configurations, the total evaluation time may extend to several days or longer.
Accordingly, the frequency, antenna configuration, Beam Direction, Operating Mode, and Power Level should be defined at the test planning stage, and the system should be capable of efficiently identifying the maximum exposure condition. IEC/IEEE 63195-1 also requires measurement reports to document the antenna tested, Operating Mode, Power Level, Frequency Band, and the rationale used to select the maximum exposure condition.
These characteristics are the reason why, in high-frequency test infrastructure, DUT control and Scanner control should not be operated as separate systems. Instead, Beam setting → output setting → Positioning → Scan → analysis should be connected within a single automated Sequence.
Spatial Averaging and Maximum Finding
Power Density compliance assessment is not performed by simply comparing the maximum electromagnetic field value at a single measurement point. Spatial-average Power Density must be calculated over an area specified by the applicable requirements from the measured spatial Field Distribution, and the maximum value across the Evaluation Surface, or psPD, must then be identified.
IEC/IEEE 63195-1 defines sPD as Power Density averaged over a specified Averaging Area and psPD as the maximum of all sPD values obtained across the Evaluation Surface. The standard identifies 1 cm² and 4 cm² as examples of Averaging Areas used depending on the applicable Exposure Limit.3 Accordingly, an arbitrary Averaging Area should not be applied. The area should be selected based on the exposure requirements and frequency conditions applicable to the product. For example, ICNIRP 2020 applies spatial averaging over 4 cm² for local Absorbed Power Density above 6 GHz and, above 30 GHz, establishes an additional limit based on a 1 cm² averaging area to account for highly concentrated exposure.4
Therefore, even when the same raw Field Data are obtained, the final result can be affected by the Sampling Grid, Reconstruction Method, Averaging Area, and Maximum Finding Algorithm. This means that data analysis software should also be validated as part of the measurement system rather than treated merely as a visualization program.
3 IEC/IEEE 63195-1:2022, Clause 3.1.2 Note 2; Annex B, Table B.5: 1 cm² and/or 4 cm² are provided as examples of averaging areas specified in exposure limits, and 1 cm² and 4 cm² psPD values are also used as system validation targets. ↩
4 ICNIRP, Guidelines for Limiting Exposure to Electromagnetic Fields (100 kHz to 300 GHz), Health Physics 118(5), 2020, Table 2 Note 5: local absorbed power density is averaged over 4 cm², and above 30 GHz, the 1 cm² averaged value is subject to an additional restriction such that it does not exceed twice the 4 cm² limit. ↩
Chamber Reflection and the Influence of the Surrounding Environment
Power Density measurement is intended to measure electromagnetic fields radiated from the DUT. In an actual laboratory, however, reflections from walls, floors, Scanner structures, DUT Holders, and surrounding metallic structures can be superimposed on the measured values.
IEC/IEEE 63195-1 manages Laboratory Reflection, RF Ambient Conditions, and Device Holder Influence as separate uncertainty contributors. The Laboratory Requirements in the standard also require measurements to be performed in a controlled environment and require management of ambient RF Noise and reflections from laboratory structures.
Therefore, designing the laboratory simply as a Shielded Room that blocks external signals is not sufficient. Absorbers should be arranged to minimize reflections around the DUT and Probe, while the Positioner and DUT Holder should also use structures and materials that minimize unnecessary scattering.
Rather than generalizing that a fully Anechoic Chamber is mandatory for every Power Density test, a more accurate approach is to design and validate the test environment so that it satisfies the RF Ambient and Reflection conditions required by the applicable measurement method.
Measurement Uncertainty and Overall System Reproducibility
Measurement uncertainty in Power Density testing cannot be represented by the Accuracy Specification of a single instrument. The uncertainty contributors identified in IEC/IEEE 63195-1 include Calibration, Probe Correction, Frequency Response, Isotropy, Linearity, Probe Positioning, Spatial Resolution, Measurement Drift, Sampling, Reconstruction, Spatial Averaging, Probe–DUT Coupling, DUT Alignment, Ambient RF, Laboratory Reflection, and DUT Drift.
In an actual test laboratory, these factors therefore need to be managed within a single Measurement Uncertainty Budget. Merely obtaining Calibration Certificates for individual Probes and measurement instruments does not automatically ensure the reliability of the overall test result.
For this reason, Calibration and System Verification need to be distinguished. Calibration is the process of determining the measurement characteristics of individual equipment, such as Probes and measurement instruments, against reference values. By contrast, System Check and System Validation confirm whether the complete measurement system, with the Scanner, Probe, Positioning, Software, and test environment combined, maintains the required accuracy and repeatability. IEC/IEEE 63195-1 also specifies Relative System Check, Absolute System Check, and System Validation as separate verification procedures.
Ultimately, the key challenge in high-frequency Power Density measurement is not to obtain high-performance Probes or Scanners individually, but to build a system capable of repeatedly reproducing the same test conditions while controlling the uncertainty associated with each measurement variable. Probe Calibration, Scanner Resolution, DUT Positioning, Beam Condition, Spatial Averaging, RF Environment, and data processing must all be managed as a single Measurement Chain. These requirements form the design criteria for the Reference Test Architecture discussed in the next section.
4. Reference Test Architecture
Test infrastructure for high-frequency Power Density measurement should not be designed around a single measurement instrument. IEC/IEEE 63195-1:2022 defines a Power Density measurement system as an integrated structure consisting of a Field Probe, Measurement Instrumentation, Scanning System, Device Holder, and Post-processing, and specifies that testing should be performed in a controlled laboratory environment. The performance of the system is also not determined solely through equipment calibration, but is verified through a Relative System Check, Absolute System Check, and System Validation.
Accordingly, when building an actual test laboratory, the Test Environment → Electromagnetic Field Measurement → Precision Positioning → DUT Control → Data Acquisition → Power Density Analysis → System Verification → Reporting and History Management process should be connected as a single Measurement Chain.
| Configuration Area | Main Components | Core Role |
|---|---|---|
| RF Test Environment | RF Shielding, Absorber, DUT/Test Zone | Control external RF interference and internal reflections |
| Power Density Measurement | E/H-field Probe, RF Instrumentation, Scanner | Measure electromagnetic field distribution over the evaluation surface |
| Positioning Architecture | DUT Holder, DUT Positioner, Probe/Scanner Positioner | Reproduce distance, position, and angle between the DUT and Probe |
| Automation & Control | Scanner Control, DUT Control, Beam/Power Setting | Implement repeatable test Sequences |
| Data Acquisition & Analysis | E/H-field Acquisition, Reconstruction, Spatial Averaging | Derive psPD and applicable assessment metrics |
| Calibration & Verification | Probe Calibration, System Check, System Validation | Verify the reliability of the entire Measurement Chain |
| Reporting & Data Management | Test Report, Configuration DB, Calibration History | Ensure traceability of test results and manage repeated testing |
4.1 RF Shielded Test Environment
Because Power Density measurement evaluates the electromagnetic field distribution around the DUT, external RF signals or reflections from surrounding structures superimposed on the measurement region can reduce the accuracy and repeatability of the results. Therefore, the test environment should be designed not merely as a physical space, but as an electromagnetic environment capable of controlling RF Ambient and Reflection.
The IEC/IEEE 63195 series requires laboratory ambient RF conditions, wall reflections, temperature conditions, and the material and reflection characteristics of the DUT Holder to be managed as part of the test conditions. IEC/IEEE 63195-2 also requires Laboratory Wall Reflection, Ambient Noise, Probe characteristics, Positioning Accuracy, and Holder Reflection to be considered for equipment used for experimental model validation.
In an actual infrastructure, an RF Shielded Environment capable of sufficiently suppressing external RF interference should first be established, while Absorbers should be arranged around the DUT and Probe as necessary to minimize unwanted reflections. The geometry and materials of the DUT Holder and Positioner should also be selected so that metallic structures do not create unnecessary scattering sources within the measurement region.
It is not necessary to generalize that a specific type of fully Anechoic Chamber must always be used for Power Density measurement. The key requirement is whether the test environment satisfies the Ambient RF, Reflection, and Scattering conditions permitted by the applicable measurement method and whether its performance can be validated.
From this perspective, the test chamber should be designed to provide the following functions:
- Shielding against external RF signals
- Control of unwanted reflections and scattering around the DUT
- Sufficient travel range for the Scanner and Positioner
- Support for various DUT Orientations and Evaluation Distances
- Minimization of interference from cables, Control Electronics, and RF measurement instruments within the measurement region
- Sufficient space for System Check and Validation using a Reference Antenna
4.2 Power Density Measurement System
The core function of a Power Density Measurement System is to measure the E-field or H-field around the DUT, or both fields when necessary, and derive the Power Density distribution over the evaluation surface.
IEC/IEEE 63195-1 specifies that the measurement system includes a Field Probe, Measurement Instrumentation, and Scanning System, and that Probe output is transferred through the measurement instrumentation to a computer for analysis. Measurement Instrumentation, as well as the Probe itself, must be calibrated, and signal-processing functions such as frequency filtering, amplification, and nonlinearity correction are also managed as part of the measurement system.
Therefore, equipment selection should consider not only the maximum frequency specification but also the performance required under actual test conditions.
The Field Probe should be evaluated not only for operating frequency range but also for Sensitivity, Dynamic Range, Linearity, Isotropy, Cross Coupling, Physical Size, Spatial Resolution, and Field Impedance Dependence. Particularly in the mmWave range, the physical size of the Probe is directly related to spatial resolution, and therefore the Probe should be selected with consideration of the Evaluation Distance and Field Gradient. IEC/IEEE 63195-1 also separately addresses characteristics such as Sensitivity, Linearity, Lower Detection Limit, Isotropy, and Response Time in the Probe Calibration process.
Measurement Instrumentation is responsible for stably acquiring the Probe output signal and transferring it to the analysis system. Depending on the configuration, this may include an RF Receiver, VNA, Signal Analyzer, or dedicated Acquisition Electronics, and calibration status should be maintained under the same interface and measurement conditions used with the Probe.
The Field Scanner is responsible for moving the Probe to the required positions within the Evaluation Region. IEC/IEEE 63195-1 requires the Scanner to support 2D or 3D Scan methods, Grid determination algorithms, Positioning Accuracy and Repeatability, Scan Resolution, and, where necessary, Probe Orientation control. A sufficient Measurement Region must also be secured so that psPD can be determined at any location on the Evaluation Surface.
Accordingly, the Power Density Measurement System should be understood as a single measurement chain:
Field Probe → Scanner → Measurement Instrumentation → Data Acquisition → Post-processing
If the accuracy of any one component is insufficient, it affects the overall Measurement Uncertainty. Equipment specifications should therefore be evaluated within the complete Measurement Chain rather than independently.
4.3 Positioning Architecture
In high-frequency Power Density measurement, the relative position between the DUT and Probe directly affects the test result. Particularly in the near field, electromagnetic field distribution can change even with distance variations on the order of several millimeters. Positioning Architecture should therefore be designed not simply as a mechanical motion system, but as a system that establishes and repeatedly reproduces the measurement coordinate system.
IEC/IEEE 63195-1 requires Scanner Positioning Accuracy and Repeatability and also treats DUT Alignment as a contributor to Measurement Uncertainty. The Absolute System Check also requires the distance between the Reference Antenna and the Evaluation Surface to be set with sufficient precision to ensure repeatability.
Positioning Architecture can generally be divided into two areas.
Probe/Scanner Positioning is the structure used to move across the Evaluation Surface according to a defined Grid and measure the Field Distribution. A Cartesian XYZ structure, multi-axis Robot, or another structure suitable for the measurement method may be used, and the configuration should be selected based on the required Scan Area and Position Accuracy.
DUT Positioning is the structure used to repeatedly place the DUT at a specified position and orientation. Translation and Rotation may be required depending on the DUT geometry and antenna configuration, and the system should be capable of changing test conditions according to Face, Edge, Orientation, or Beam Direction.
The important point is not to standardize the Positioner according to a specific number of axes or mechanical configuration, but to design the Positioning Architecture according to the test objective, DUT geometry, and measurement method.
For example, in a simple planar scan, a structure in which the DUT remains fixed while only the Probe moves may be efficient. For products requiring evaluation of multiple faces or various Beam Directions, however, a combined structure using DUT Rotation and Scanner Translation may be more appropriate.
4.4 Automation & Control
Because 5G NR FR2 devices use Array Antennas and Beamforming, a single product may require evaluation of multiple Antenna Configurations and Beam States. Test automation therefore becomes a key factor determining overall test time and result repeatability rather than merely a convenience feature.
Studies of Power Density assessment for actual 5G mmWave handsets have noted that measuring each Beam Realization using conventional Planar Near-field methods can significantly increase total test time. Accordingly, fast scanning techniques and Reconstruction-based assessment methods using Intermediate-field data have been proposed.
At a minimum, the test automation system should be capable of connecting the following elements:
DUT Configuration → Beam Setting → RF Output Setting → Positioning → Scan → Data Acquisition → Analysis
If only the Scanner and Positioner are automated while the DUT Beam or Output Mode is changed manually by the operator, operational errors and data omissions may occur during repeated testing. Where possible, it is therefore desirable to integrate the DUT Control Interface with the measurement system so that Antenna Configuration, Channel, Beam, Modulation, and Output Condition can be managed within the same Sequence.
The test reporting requirements of IEC/IEEE 63195-1 also require documentation of the tested Frequency Band, Modulation, Position, Orientation, Antenna Configuration, Operating Mode, and Power Level, together with the rationale for selecting the maximum exposure Configuration.
Automation also serves to program these test conditions so that the same Sequence is executed when the same test is repeated. This not only improves test speed but also reduces operator dependency and improves Repeatability.
4.5 Data Acquisition & Power Density Analysis
Field Data collected by the Scanner is not itself the final compliance assessment result. The measured data must be converted into the final Power Density result through correction, Reconstruction, and Spatial Averaging as required.
IEC/IEEE 63195-1 and IEC/IEEE 63195-2 use a process in which Spatial-average Power Density is derived from measured or computed Field Data and Peak Spatial-average Power Density is then identified across the Evaluation Surface. Averaging Areas such as 1 cm² or 4 cm² may be applied, and related Post-processing methods are commonly used in both standards.
The Data Analysis System may generally include the following functions:
- E-field/H-field Data Acquisition
- Application of Probe Calibration Data
- Probe Correction
- Measurement Grid generation
- Field Reconstruction where required
- Incident Power Density calculation
- Spatial Averaging
- Peak Spatial-average Power Density search
- DUT Output Drift Correction
- Incorporation of Measurement Uncertainty
- Result Map and Visualization generation
Particularly for measurement methods in which the evaluation surface does not coincide with the actual Scan Plane, the accuracy and Validation of the Reconstruction Algorithm become important. IEC/IEEE 63195-2 also specifies that Computational Software should comply with and be validated against the prescribed methods for determining sPD and psPD.
The analysis software should therefore be managed not as an auxiliary program that merely generates graphs, but as part of the Measurement System that determines the compliance assessment result.
4.6 System Verification & Calibration
One concept that must be clearly distinguished when building test infrastructure is the difference between Calibration and System Validation.
Calibration is the process of comparing the measurement characteristics of individual instruments such as the Probe and Measurement Instrumentation against reference values. System Verification and Validation, by contrast, determine whether the complete Measurement System, including the Probe, Scanner, Positioner, RF Instrumentation, Software, and test environment, performs the actual Power Density measurement correctly.
IEC/IEEE 63195-1 Annex A divides this process into the following three stages.
- Relative System Check
- Determine whether system performance has changed relative to an established reference condition
- Absolute System Check
- Verify whether the installed system accurately measures a known Target Value
- System Validation
- Comprehensively validate measurement performance and uncertainty across the full applicable range of the system
The Absolute System Check uses a calibrated Reference Antenna and applies the same System Components and Software Settings used in actual compliance testing. The measured psPD is then compared with a Target Value. This allows installation errors or degradation in system performance to be identified.
System Validation is performed under a broader range of conditions. IEC/IEEE 63195-1 requires multiple Validation Antennas, different Power Levels, and Communication Signals to be used to verify the Dynamic Range and Modulation Response. Examples in the standard include not only CW but also QPSK, 16-QAM, 64-QAM, and 5G NR CP-OFDM and DFT-s-OFDM signals as Validation conditions.
Accordingly, an actual test laboratory should establish a quality-management process that follows the sequence:
Probe Calibration → System Check → System Validation → Routine Test
4.7 Reporting & Data Management
In high-frequency Power Density testing, storing only a single measurement value is insufficient to reproduce the same result. The DUT Antenna Configuration, Beam State, Channel, Output Power, Evaluation Distance, Position, Orientation, Probe, Calibration Data, Software Version, and System Check status must all be linked to the result.
IEC/IEEE 63195-1 requires the test report to include the DUT geometry, test position and orientation, Antenna Configuration, Operating Mode, Power Level, Frequency Band, rationale for selecting the maximum exposure condition, psPD result, Scan result, and relevant Setup Parameters. The Validation procedure for the Measurement Method and the PD Uncertainty must also be documented.
The Reporting System should therefore be designed to ensure Traceability of test data.
At a minimum, the test database should preferably connect the following information:
DUT ID → Hardware/Software Version → Test Configuration → Beam/Channel → Position → Measurement Data → Analysis Result → Uncertainty → Calibration/Validation Status
This structure enables systematic retesting of the same product, comparison before and after design changes, analysis of the impact of calibration changes, and tracking of test results.
Automated Report Generation can also reduce omissions or input errors that may occur when operators manually transfer results. Particularly for 5G and 6G products with large numbers of Beams and Operating Modes, a structure that automatically links Raw Data and Final Results can significantly improve test efficiency.
In summary, the core of the Reference Test Architecture is not the combination of individual measurement instruments, but the integration of the entire test workflow. Within an RF Shielded Test Environment, the DUT must be positioned precisely, the Scanner and Probe must measure the Field Distribution, the DUT Beam and Output Condition must be controlled automatically, and the measured data must be processed through Spatial Averaging and Uncertainty analysis before being connected to the psPD result and final Report.
Test infrastructure designed in this manner can support not only current Power Density compliance assessment for 5G NR FR2, but also future changes in frequency bands and Antenna Architectures by allowing Probes, Positioners, RF Instrumentation, and Software to be replaced or expanded in stages. In other words, the objective of laboratory construction is not to create one fixed measurement system, but to establish modular and scalable Test Infrastructure capable of adapting to changing measurement requirements.
5. Implementation Guide
Building high-frequency Power Density test infrastructure should not begin by selecting a chamber and measurement equipment independently and then combining them. Instead, the appropriate approach is to first define the DUT and applicable standards, and then work backward from those requirements to design the test environment, measurement, Positioning, verification, and automation systems.
In particular, the measurement procedures in IEC/IEEE 63195-1:2022 address not only the performance of the Field Probe or Scanner, but also the Measurement Region, Evaluation Distance, DUT Positioning, Laboratory Reflection, System Check, System Validation, and Measurement Uncertainty within a single measurement framework. If the overall Measurement Chain is not defined during the initial design stage, problems such as insufficient measurement area, inadequate Positioning Accuracy, difficulties in system validation, or a lack of automation interfaces may arise even after the equipment has been installed.
The process of building a test laboratory can be organized into the following steps.
| Implementation Step | Key Considerations | Main Deliverable |
|---|---|---|
| Step 1. Define the Scope |
DUT, frequency, applicable standards, assessment metrics, measurement method | Test Requirement |
| Step 2. Design the Test Environment |
RF Shielding, Reflection, Absorber, space and workflow | Chamber/Test Environment |
| Step 3. Build the Measurement System |
Probe, Scanner, RF Instrumentation | Measurement Chain |
| Step 4. Build the Positioning System |
Evaluation Distance, Position Accuracy, DUT Alignment | Positioning Architecture |
| Step 5. Calibration & Verification |
Calibration, System Check, System Validation | Measurement Reliability |
| Step 6. Automation Integration |
Integration of DUT, Beam, Scanner, Positioner, and instrumentation | Automated Test Sequence |
| Step 7. Data & Reporting |
Analysis, uncertainty, reporting, history management | Test Database & Report |
| Step 8. Operation & Upgrade |
Expansion for new frequencies, Probes, DUTs, and software | Future-ready Infrastructure |
Step 1. Define the Test Objective and Scope
The first step in building test infrastructure is not deciding which equipment to purchase, but defining what kind of DUT will be evaluated and under what conditions.
IEC/IEEE 63195-1:2022 covers Power Density measurement procedures for wireless devices operating in close proximity to the human head or body over the frequency range of 6 GHz to 300 GHz. By contrast, IEC/IEEE 62209-1528:2020 specifies SAR measurement procedures for hand-held and body-mounted wireless devices operating from 4 MHz to 10 GHz. Therefore, when defining the scope of a laboratory, it is more appropriate to base it on the target product, frequency range, applicable exposure metric, and test procedure rather than simply describing it as a “5G laboratory” or “mmWave laboratory.”
For a wireless device operating in close proximity, such as a smartphone, the Frequency Bands supported by the product, the number and location of Antenna Modules, Beamforming capability, actual use distance, and Orientation should first be reviewed. The applicable regulations and standards can then be used to determine the required SAR or Power Density assessment scope and Measurement Method.
At this stage, it is important to define the following items in detail:
- Target DUTs and product categories
- Frequency range to be tested
- DUT Antenna Configuration and Beamforming characteristics
- Use position and Evaluation Distance
- Applicable Exposure Metric
- Applicable international standards and regulatory requirements
- Scope of Measurement or Computational Procedures
- Required Scan Area and Dynamic Range
- Future frequency bands and product categories to be supported
IEC/IEEE 63195-1 and IEC/IEEE 63195-2 both cover the same frequency range of 6 GHz to 300 GHz, but the former specifies the Measurement Procedure while the latter specifies the Computational Procedure. Therefore, when building a physical measurement laboratory, system requirements should be defined primarily around IEC/IEEE 63195-1. If computational or Simulation-based assessment will also be performed, a separate Computational Validation framework based on IEC/IEEE 63195-2 should also be considered.
Step 2. Design the Test Environment
Once the Scope has been defined, the next step is to design an RF Test Environment in which the measurement system can operate reliably.
In Power Density measurement, not only external RF signals but also reflections and scattering from chamber walls, floors, DUT Holders, Positioners, and surrounding structures can affect the measurement results. The test environment should therefore be designed to shield external interference and control internal reflections to the required level.
At this stage, reviewing RF Shielding performance alone is not sufficient. The actual usable space inside the chamber should be determined by considering the relative position of the DUT and Evaluation Surface, the maximum travel range of the Scanner, the rotational envelope of the DUT Positioner, and the space required for System Check using a Reference Antenna.
In particular, chamber size should not be determined solely by the physical size of the DUT.
DUT installation space + Evaluation Distance + Scan Area + Scanner Stroke + Positioner operating range + clearance from surrounding structures must all be considered together.
Absorbers should also not simply be applied uniformly throughout the chamber. Their arrangement should be determined based on Reflection and Scattering that may occur around the DUT and Measurement Region. Because IEC/IEEE 63195-1 treats Laboratory Reflection and RF Ambient Conditions as contributors to measurement uncertainty, it should be possible to verify after installation that these effects are sufficiently controlled under actual measurement conditions.
For this reason, it is important to review the requirements of the measurement system and Positioner together from the chamber design stage. If the chamber is completed first and the Scanner and Positioner are installed afterward, the actual measurable region may be restricted by insufficient travel range or structural interference.
Step 3. Build the Measurement System
Once the test environment has been defined, the Probe, Scanner, and Measurement Instrumentation can be configured.
The key objective at this stage is not to maximize the specifications of individual equipment, but to configure a Measurement Chain that satisfies the test Scope defined in Step 1.
For example, a Field Probe should be evaluated not only in terms of its maximum operating frequency but also its Sensitivity, Dynamic Range, Linearity, Isotropy, Spatial Resolution, Cross Coupling, and Calibration conditions. At high frequencies, the physical size of the Probe can affect spatial resolution, so the operating frequency and Field Gradient should also be considered.
The Scanner should be capable of covering the required Measurement Region while providing the necessary Sampling Grid and Position Accuracy. Even if the Probe itself has sufficient spatial resolution, it is difficult to ensure overall measurement accuracy if the Scanner lacks adequate Position Accuracy and Repeatability.
RF Measurement Instrumentation should also be selected together with the Probe rather than independently, based on the Frequency, Bandwidth, Dynamic Range, and required Amplitude/Phase information of the signal.
Accordingly, the specifications of the Measurement System should be determined according to the following relationship:
Test frequency and DUT → required spatial resolution → Probe → Scanner → RF Instrumentation → Acquisition Software
Designing the system in this manner helps prevent the performance of individual equipment from being either excessive or insufficient while allowing the overall Measurement Uncertainty to be managed effectively.
Step 4. Build the Positioning System
Because the relative position between the DUT and Probe directly affects Power Density measurement results, the Positioning System should not be designed independently of the Measurement System.
The Positioning Architecture can broadly be divided into Probe/Scanner Positioning and DUT Positioning.
Probe Positioning should provide sufficient travel range and Position Accuracy to measure the Field Distribution over the Evaluation Surface according to the specified Grid. DUT Positioning should be capable of repeatedly reproducing the position and angle of the DUT according to the Surface, Orientation, Antenna Module, and Beam Direction to be tested.
IEC/IEEE 63195-1 treats Probe Positioning Offset and Repeatability, as well as DUT Alignment, as contributors to Measurement Uncertainty. Therefore, in an actual system, it is necessary to verify not only the nominal positioning accuracy of the motor and Stage, but the relative coordinate accuracy of the entire DUT–Probe–Evaluation Surface configuration.
The DUT Fixture is also an important component. Even after the same product has been removed and reinstalled, the same position and Orientation should be reproducible, while the influence of the Holder itself on the RF Field should be minimized.
Rather than applying a single mechanical configuration uniformly, the degree of freedom required for the Positioning Architecture should be determined according to the test objective. A relatively simple Cartesian Scanner may be suitable for evaluating a single Surface, while a multi-axis configuration incorporating a DUT Rotation Axis may be more efficient when multiple Orientations or Antenna Modules need to be evaluated automatically.
Step 5. Establish a Calibration & System Verification Framework
The installation of measurement equipment does not mean that the test system is immediately ready for compliance assessment.
The Calibration status of individual measurement instruments, including the Probe and Measurement Instrumentation, should first be confirmed. Probe Calibration should verify not only Sensitivity over the operating frequency range, but also whether Linearity, Isotropy, and other measurement characteristics satisfy the test requirements.
Calibration alone, however, cannot confirm the accuracy of the complete system. The entire Measurement Chain must be verified with the Scanner Positioning, DUT Holder, Measurement Software, and test environment combined.
IEC/IEEE 63195-1 distinguishes Relative System Check, Absolute System Check, and System Validation for this purpose.
| Quality Control Stage | Main Purpose | Typical Timing |
|---|---|---|
| Calibration | Verify measurement characteristics of the Probe and instrumentation | Periodic calibration and as required |
| Relative System Check | Identify changes in system condition relative to an established baseline | Routine Check |
| Absolute System Check | Verify overall system performance against a known reference value | Installation, modification, and periodic inspection |
| System Validation | Validate measurement method and system performance over the applicable range | Initial installation and major system changes |
Reference Sources or Reference Antennas, Fixtures, and predefined Measurement Sequences required for System Check and Validation should be prepared from the test laboratory design stage.
In other words, building a laboratory means establishing not only measurement equipment but also a verification framework capable of continuously demonstrating that the equipment is operating correctly.
Step 6. Automation & Software Integration
High-frequency wireless devices may require repeated evaluation across multiple Frequencies, Channels, Antenna Configurations, and Beam Conditions. A manual test process can therefore significantly increase the total test time and operator dependency.
The Scanner, Positioner, DUT, and RF Instrumentation should therefore be integrated into a single Control Software environment wherever possible.
An automated Sequence can, for example, be configured as follows:
DUT Identification → Frequency/Channel Setting → Antenna/Beam Setting → Output Verification → Positioning → Scan → Data Acquisition → Spatial Averaging → Maximum Finding → Result Storage
The purpose of automation is not merely to reduce measurement time. More importantly, it improves the reproducibility of test conditions by repeatedly applying the same Position, Beam, Output, and Scan Conditions while reducing Human Error that may occur during operation.
Because DUT Output Drift may occur during testing, the Sequence may also include Reference Measurements before and after the test to identify or compensate for changes in output where necessary.
Software design should also consider the future addition of new DUT Control Interfaces or RF Instrumentation. A modular architecture in which device-specific Drivers or Interfaces can be replaced is more suitable for long-term laboratory operation than a closed structure that is tightly coupled to specific equipment.
Step 7. Build the Data & Reporting System
Power Density testing requires traceability not only of the measurement result, but also of the conditions under which that result was produced.
A data structure is therefore required that connects Raw Measurement Data with DUT Configuration, Position, Beam Condition, Calibration Data, System Check results, Analysis Parameters, and Measurement Uncertainty under the same test ID.
Test data can be managed using a structure such as:
Project → DUT → Test Configuration → Frequency/Beam → Position → Raw Data → Analysis → Uncertainty → Final Result
The reporting requirements of IEC/IEEE 63195-1 also require documentation of information necessary to reproduce the result, including DUT test position and Orientation, Frequency Band, Antenna Configuration, Operating Mode, and Power Level.
The Reporting System can be configured to automatically retrieve these data and populate a predefined Template with the Power Density Map, Peak Spatial-average Power Density, test conditions, applicable Calibration information, and uncertainty results.
This not only reduces errors that may occur during manual data entry but also facilitates retesting of the same DUT and comparison of results between product Revisions.
Step 8. Operation & Upgrade Strategy
The final step is to establish an operation and expansion strategy for the laboratory after construction.
If Power Density test infrastructure is optimized only for products available at the time of installation, major reconstruction may be required to accommodate new Frequency Bands or Antenna Architectures in the future. Because the specific frequencies and compliance assessment frameworks for future 6G systems have not yet been fully established, it is more important at this stage to secure an expandable Architecture than to fix the system around a particular assumed 6G test condition.
The test environment and system can therefore be modularized wherever practical.
The Probe can be configured as a replaceable component to accommodate new frequency ranges or Spatial Resolution requirements. The Scanner and Positioner can be designed with sufficient travel range and interfaces to support future changes in DUT size and test Orientation. RF Instrumentation can likewise be configured so that it can be replaced or expanded as required frequency and bandwidth needs evolve.
The same principle applies to Software. When a new Probe, Positioner, or RF Instrument is added, a system that allows a new device Interface or Analysis Module to be added is more efficient over the long term than one that requires the entire program to be redeveloped.
From this perspective, laboratory scalability can be divided into four areas.
- Frequency Scalability
- Ability to replace or add Probes and RF Instrumentation
- Mechanical Scalability
- Expansion of Scanner Stroke, DUT size, and Positioner Axes
- Software Scalability
- Addition of new equipment Drivers, Measurement Methods, and Analysis Modules
- Data Scalability
- Database structure capable of accommodating new test items and DUT Configurations
In summary, building Power Density test infrastructure does not end with installing the equipment currently required. The test Scope should first be defined, and the RF Test Environment, Measurement System, Positioning Architecture, Calibration and Validation, Automation, and Data Management should then be developed as one integrated workflow based on those requirements.
In particular, rather than selecting the chamber, Scanner, Probe, Positioner, and RF Instrumentation independently, an All-in-One Engineering approach should be applied from the initial design stage to review the spatial, mechanical, electromagnetic, and software interfaces of the entire system together. This helps reduce equipment interference, insufficient measurement area, and automation compatibility issues that may otherwise arise after installation, while improving the accuracy and repeatability of test results.
Ultimately, this implementation strategy provides the foundation for sustainable high-frequency human exposure assessment infrastructure that can support current Power Density testing for 5G NR FR2 and can also be expanded in stages as frequencies, product architectures, and assessment requirements evolve in 5G-Advanced and future 6G systems.
6. Conclusion
Human exposure assessment in the 5G and 6G era is not a process of simply replacing the existing SAR-based assessment framework with Power Density measurement. IEC/IEEE 62209-1528:2020 addresses SAR measurement procedures for hand-held and body-mounted wireless devices in the frequency range of 4 MHz to 10 GHz, while IEC/IEEE 63195-1:2022 and IEC/IEEE 63195-2:2022 specify Power Density measurement and computational procedures, respectively, for wireless devices operating in close proximity over the frequency range of 6 GHz to 300 GHz. Test laboratories and manufacturers must therefore select the appropriate Exposure Metric and test method according to the product frequency, conditions of use, applicable regulations, and assessment objectives.
Power Density becomes increasingly important at high frequencies because the penetration depth of RF energy into human tissue decreases and absorption becomes concentrated near the body surface. ICNIRP 2020 and IEEE C95.1-2019 have reorganized the framework for applying SAR and Power Density-related metrics around the 6 GHz transition region, and related studies have also emphasized the importance of new Dosimetric Metrics for evaluating near-surface energy absorption and thermal effects at frequencies above 6 GHz.
In actual compliance testing, however, the greater challenge is not the introduction of a new assessment metric itself, but implementing an accurate and repeatable measurement environment at high frequencies. In the mmWave range, wavelengths are short and near-field distributions are complex. As a result, Probe Spatial Resolution, Scanner Position Accuracy, the distance between the DUT and Evaluation Surface, DUT Alignment, Beam Direction, Output Condition, Spatial Averaging, and Laboratory Reflection can all affect the final result.
In particular, 5G NR FR2 devices using array antennas and Beamforming are not systems in which only a single fixed Radiation Pattern needs to be evaluated. The spatial electromagnetic field distribution can vary according to the Beam Configuration, and when multiple Beam Realizations must be assessed, test time using conventional Planar Near-field Scan methods can increase significantly. Studies have identified repeated evaluation of multiple Beams and measurement directions as a major efficiency challenge in Power Density compliance testing, and methods such as Field Reconstruction and the Equivalent Currents Method are being investigated to improve this process.
Accordingly, a high-frequency Power Density laboratory should not be designed as a collection of independently installed Probes, Scanners, or RF Instruments. It should instead be structured as a single integrated system:
RF Test Environment → Measurement System → Positioning Architecture → DUT/Beam Control → Data Acquisition → Power Density Analysis → Calibration & System Validation → Reporting & Data Management
Within this structure, each element is interconnected rather than independent. For example, Scanner Position Accuracy is directly related to the reproducibility of the Evaluation Distance, while Probe Spatial Resolution affects the Scan Grid and Measurement Time. The DUT Beam Control method is linked to the Automation Sequence and data structure, and the influence of Chamber Reflection and the DUT Holder must be incorporated into Measurement Uncertainty.
The reliability of test results is not determined by a single high-performance measurement instrument. This is also why IEC/IEEE 63195-1 specifies Relative System Check, Absolute System Check, and System Validation separately from Probe Calibration. The entire test system must be continuously verified under reference conditions so that reproducibility and traceability can be maintained during actual compliance testing.
Test infrastructure should also be designed to accommodate future changes rather than being fixed solely around current 5G NR FR2 requirements. Because the specific frequency use and compliance assessment framework for 6G have not yet been fully established, it is more practical at this stage to secure a modular architecture in which Probes, RF Instrumentation, Positioners, Scanners, and Software can be replaced or expanded in stages rather than treating specific candidate technologies as if they were already current regulatory requirements.
From the perspective of test laboratories and manufacturers, long-term competitiveness depends not on how high the specifications of a single test instrument are, but on whether the test Scope can be expanded in line with changing products and standards while maintaining the same level of measurement quality.
Accordingly, the core principle of human exposure assessment infrastructure in the 5G and 6G era can be summarized in one sentence:
The accuracy and reproducibility of Power Density testing are secured not by individual measurement instruments, but by an integrated Measurement Infrastructure that connects the test environment, measurement, Positioning, verification, automation, and data management.
By establishing this type of test infrastructure in advance, product development verification, certification support, repeated testing, design-change comparison, and quality management can all be operated within a single test framework, while also providing greater flexibility in responding to future changes in high-frequency wireless technologies and assessment requirements.
7. Appendix
7.1 Major International Standards and Guidelines
The major international standards and guidelines related to human exposure assessment and Power Density measurement discussed in this white paper are summarized below.
| Standard / Guideline | Scope | Role in This White Paper |
|---|---|---|
| IEC/IEEE 62209-1528:2020 | 4 MHz to 10 GHz hand-held and body-mounted wireless devices | SAR measurement procedures, human models, instrumentation, and system requirements |
| IEC/IEEE 63195-1:2022 | 6 GHz to 300 GHz wireless devices operating in close proximity | Power Density Measurement Procedure |
| IEC/IEEE 63195-2:2022 | 6 GHz to 300 GHz wireless devices operating in close proximity | Power Density Computational Procedure |
| ICNIRP Guidelines 2020 | 100 kHz to 300 GHz | RF human exposure limits and Dosimetric criteria |
| IEEE Std C95.1-2019 | 0 Hz to 300 GHz | EMF human exposure Safety Limits, including RF |
| IEEE Std C95.3-2021 | 0 Hz to 300 GHz | Recommended Practice for EMF Exposure Measurement and Computation |
IEC/IEEE 62209-1528:2020 specifies SAR measurement procedures for hand-held and body-mounted wireless devices over the frequency range of 4 MHz to 10 GHz.
IEC/IEEE 63195-1:2022 specifies Power Density measurement procedures for wireless devices used in close proximity to the head and body over the frequency range of 6 GHz to 300 GHz, while IEC/IEEE 63195-2:2022 specifies computational procedures for the same scope.
The ICNIRP 2020 Guidelines provide a framework for RF EMF human exposure limits over the frequency range of 100 kHz to 300 GHz, while IEEE C95.1-2019 specifies human exposure Safety Levels over the frequency range of 0 Hz to 300 GHz.
IEEE C95.3-2021 provides Best Practices for the measurement and computation of Electric, Magnetic, and Electromagnetic Fields related to human exposure over the frequency range of 0 Hz to 300 GHz.
7.2 Key Terms
- SAR
- Specific Absorption Rate. RF power absorbed by human tissue per unit mass
- Power Density
- Power passing through or absorbed over a unit area
- IPD
- Incident Power Density. Power density of an electromagnetic field incident on the human body surface
- APD
- Absorbed Power Density. Power density absorbed into tissue through the human body surface
- sPD
- Spatial-average Power Density. Power Density spatially averaged over a specified area
- psPD
- Peak Spatial-average Power Density. Maximum sPD obtained over the Evaluation Surface
- DUT
- Device Under Test. Equipment being tested
- Evaluation Surface
- Reference surface used for Power Density compliance assessment
- Evaluation Distance
- Distance between the DUT and the Evaluation Surface
- Probe
- Sensor used to measure electromagnetic field components such as E-field or H-field
- Scanner
- System that moves the Probe to specified measurement positions
- Positioner
- Mechanical system that controls the position and angle of the DUT or Probe
- Beamforming
- Technology that controls the phase and amplitude of an array antenna to concentrate RF energy in a specific direction
- Measurement Uncertainty
- Quantitative evaluation of uncertainty associated with a measured value
- Calibration
- Process of verifying the measurement characteristics of an instrument against a reference value
- System Check
- Check used to identify changes or abnormalities in the overall system
- System Validation
- Comprehensive procedure used to verify that the Measurement System satisfies required accuracy and performance
At RF frequencies above 6 GHz, the Power Penetration Depth in human tissue becomes shorter and near-surface absorption becomes increasingly important. Power Density-related metrics are therefore used as important Dosimetric Metrics. Related reviews also explain the rationale for the adoption of Absorbed/Epithelial Power Density concepts above 6 GHz in ICNIRP 2020 and IEEE C95.1-2019.
7.3 Power Density Test Infrastructure Implementation Checklist
Before constructing an actual test laboratory, it is advisable to review the following items in advance.
- Test Scope
- DUT type, Frequency Range, applicable standards, Exposure Metric, Evaluation Distance
- RF Environment
- Shielding, Ambient RF, Reflection, Absorber, Test Zone size
- Probe
- Frequency Range, Sensitivity, Dynamic Range, Isotropy, Spatial Resolution, Calibration
- Scanner
- Scan Area, Grid Resolution, Position Accuracy, Repeatability
- DUT Positioning
- Orientation, Rotation, Evaluation Distance, Alignment repeatability
- RF Instrumentation
- Frequency/Bandwidth, Dynamic Range, Amplitude/Phase Measurement
- DUT Control
- Frequency, Channel, Antenna, Beam, Output Mode control
- Automation
- Integrated control of Scanner, Positioner, DUT, and measurement instruments
- Data Analysis
- Field Correction, Reconstruction, Spatial Averaging, psPD search
- Uncertainty
- Budget for Probe, Positioning, Sampling, Reflection, DUT Drift, and related factors
- System Verification
- Relative Check, Absolute Check, System Validation
- Reporting
- Automatic recording of test conditions, DUT Configuration, Calibration, Uncertainty, and results
- Data Management
- Tracking of Raw Data, Analysis Results, Validation History, and Software Version
- Upgrade Strategy
- Ability to add new Probes, RF Instruments, Positioners, and Software Modules
7.4 Major Components of Measurement Uncertainty
Measurement Uncertainty in IEC/IEEE 63195-1 is not determined by Probe Accuracy alone. In an actual measurement system, multiple factors must be considered simultaneously.
Measurement-related factors include Probe Calibration, Frequency Response, Isotropy, Linearity, Sensor Cross Coupling, Probe Correction, and errors associated with Measurement Instrumentation.
Positioning-related factors include Probe Positioning Offset, Positioning Repeatability, DUT Alignment, and errors in setting the Evaluation Distance.
Sampling and analysis-related factors may include uncertainties arising from Spatial Resolution, Grid Sampling, Field Reconstruction, Spatial Averaging, and Maximum Finding.
Test environment-related factors include Laboratory Reflection, Ambient RF Conditions, the influence of the DUT Holder, and scattering from surrounding structures.
Finally, DUT-related variables such as Output Drift, Operating Mode, and the repeatability of Beam Conditions can also affect the overall result.
Accordingly, the Measurement Uncertainty Budget should not be designed as a simple sum of individual equipment specifications. It should instead identify and quantify uncertainty contributors across the entire actual test Sequence.
7.5 Summary of the Reference Test Infrastructure
The Reference Test Infrastructure proposed in this white paper can be summarized as the following workflow.
- 1. RF Shielded Test Environment
- Suppression of external RF interference and control of reflections and scattering within the Measurement Region
- 2. Power Density Measurement System
- Measurement of Field Distribution using a Probe, Scanner, and RF Measurement Instrumentation
- 3. Positioning Architecture
- Precise control of the Evaluation Distance, Position, and Orientation between the DUT and Probe
- 4. Automation & Control
- Automated Sequence control of Beam, Output Condition, Scanner, Positioner, and RF Instrumentation
- 5. Data Acquisition & Analysis
- Field Data acquisition, required correction and Reconstruction, Spatial Averaging, and psPD analysis
- 6. Calibration & System Verification
- Calibration of individual instruments and Check and Validation of the complete Measurement System
- 7. Reporting & Data Management
- Management of DUT Configuration, test conditions, measurement results, Uncertainty, and Calibration/Validation history
The core principle of this structure is to treat each component not as an independent piece of equipment, but as part of a single integrated system designed to ensure Traceability and Repeatability of test results.
7.6 Final Questions Before Implementation
Before detailed laboratory design and equipment procurement, the following questions should be answered clearly.
- What products will be tested?
- The size, usage conditions, Antenna Architecture, and Beamforming structure of the target DUT should be defined.
- What frequency range should the laboratory support?
- The required future Frequency Expansion range should be determined based not only on current products but also on actual business plans.
- Which Exposure Metrics will be evaluated?
- Applicable metrics such as SAR, IPD, and APD should be distinguished according to the relevant regulations and standards.
- How will the Evaluation Surface be implemented?
- The distance from the DUT, Scan Area, Sampling Resolution, and whether Reconstruction will be used should be determined.
- How will the maximum exposure condition be identified?
- The search method for Frequency, Channel, Antenna, Beam, Orientation, and Output Condition should be defined.
- How will Measurement Uncertainty be verified?
- The verification framework should include not only Probe Calibration, but also Positioning, Reflection, Sampling, and overall System Validation.
- How will repeated testing be automated?
- It should be confirmed whether not only the Scanner and Positioner, but also DUT Configuration and Measurement Instrumentation can be connected within a single Sequence.
- How will the system be expanded in the future?
- Mechanical, RF, and Software Interfaces should be designed with sufficient scalability to support the addition of new Probes, measurement instruments, Positioners, and Software.
Addressing these questions during the design stage allows the laboratory project to move beyond simple equipment procurement and become a long-term High-Frequency RF Human Exposure Compliance Test Infrastructure implementation project.